3 Phase - Line Impedance Stabilization Network

Model: LI-3P-1X

150 kHz to 30 MHz

LI-550A

3 Phase LISN for CISPR / ANSI

The LI-3P-1x series consists of four separate models of four-conductor, 50Ω / 50 μH Line Impedance Stabilization Networks (LISNs). The primary differences between the four models are their respective current ratings:

LI-3P-116 16 Amps (per line, continuous)
LI-3P-132 32 Amps (per line, continuous)
LI-3P-163 63 Amps (per line, continuous)
LI-3P-1100 100 Amps (per line, continuous)

The side panels for each LISN are louvered for cooling purposes. The LI-3P-163 and LI-3P-1100 also include two internal cooling fans operated by a switch on the rear panel.

Three phase LISNs

can be operated remotely or locally. Remote switching of the line under test (L1, L2, L3, N) is performed using the RLI-100 Remote LISN Interface, which controls the LISN via fiber optic connection.
In addition to the remote method, the line under test can also be selected using the mechanical, four-position switch located on the front panel of the LISNs. Using either switching method, the lines which are not selected are internally terminated into 50 ohms, while the selected line is terminated by the 50 ohm input impedance of the measuring instrument.

  • Frequency Range: 150 kHz to 30 MHz
  • Current Ratings of 16, 32, 63 and 100 Amps
  • Fully Compliant with CISPR 16-1-2/ANSI C63.4
  • Remote Switching of Line Under Test
  • Four-conductor, 50Ω / 50 μH Networks appropriate for both 3 Delta and Wye Power Configurations
  • Three-Year Warranty


  • LI-3P-116 LI-3P-132 LI-3P-163 LI-3P-1100
    GENERAL
    Products Description Line Impedance Stabilization Network (LISN)
    Application Power Line Conducted Emissions Tests
    Standards ANSI C63.4, CISPR 16-1-2
    Type 50Ω / 50 μH, (4) Conductor Network
    Frequency Range 150 kHz to 30 MHz
    Insertion Loss (Voltage Division Factor) <0.7 dB
    Isolation >40 dB
    INPUT POWER RATINGS FOR EQUIPMENT UNDER TEST (EUT)
    Current (maximum continuous, per line) 16 Amperes 32 Amperes 63 Amperes 100 Amperes
    AC Voltage (maximum) 500 Voltsrms (line to line), 288 Voltsrms (line to ground)
    DC Voltage (maximum) 705 Volts DC
    ELECTRICAL
    Remote Interface Power Inputs 6 Volts DC (unregulated), 500 mA (LISN and RLI-100 Remote LISN Interface)
    Cooling Fans Power Input Not Applicable 15 Volts DC (unregulated), 500 mA
    INPUT/OUTPUT CONNECTORS
    Power Input Port Plug (affixed to LISN chassis) Schneider Electric P/N: 83862 Schneider Electric P/N: 83874 Schneider Electric P/N: 81886 Schneider Electric P/N: 81898
    Power Input Socket (for power input cable) Schneider Electric P/N: PKF16M745 Schneider Electric P/N: PKF32M745 Schneider Electric P/N: 81486 Schneider Electric P/N: 81498
    Power Output Port Socket (affixed to LISN chassis) Schneider Electric P/N: PKF16F745 Schneider Electric P/N: PKF32F745 Schneider Electric P/N: 81286 Schneider Electric P/N: 81298
    Power Output Port Plug (for EUT power cable) Schneider Electric P/N: PKE16M745 Schneider Electric P/N: PKE32M745 Schneider Electric P/N: 81386 Schneider Electric P/N: 81398
    RF Measurement Port 50Ω - N-Type (female)
    Fiber Optic Ports Avago Duplex Latching POF Jack (LISN and RLI-100 Remote LISN Interface)
    Remote Interface Power Input Ports 5.5/2.5 mm Power Jack (LISN and RLI-100 Remote LISN Interface)
    Cooling Fans Power Input Port Not Applicable 5.5/2.1 mm Power Jack
    ENVIRONMENTAL
    Operating Temperature 40F to 104F (5C to 40C)
    Cooling Louvered Side Panels Louvered Side Panels
    Forced Air by (2) user-controlled, internal fans with (2) 4.5 circular intake openings on rear panel (each opening protected by a circular metal finger guard)
    (2) 4 square air outlets located on the top cover (each opening protected by metallic mesh)
    Dimensions & Weight See Datasheet

    Typical Impedance Data

    Imepdance

    Typical Phase Data

    Imepdance

    Typical Insertion Loss

    Imepdance

    Typical Isolation Data

    Imepdance