A 100 Watts Conducted Immunity System (High Power)

CIS-100

150 kHz to 230 MHz

conducted immunity test system

A 100 Watts Conducted Immunity System (High Power)

  • Conducted Immunity System (CIS) to Test According to IEC 61000-4-6. CIS-100 is capable of generating power upto 100 watts. 
  • The CIS series test systems are configured to perform conducted immunity testing according to IEC 61000-4-6.
  • The system includes ACS series power amplifier, Coupling Decoupling Network (CDN), directional couplers, attenuators, cables and CSAT software.
  • These systems when properly configured are capable of testing your products to the all levels given in IEC 61000-4-6 up to 230 MHz.
  • The IEC-61000-4-6 requires that an appropriate a coupling decoupling networks (CDN) with defined common mode impedance is used for testing conducted immunity. Com-Power has a wide variety of CDNs. The system can be customized include the CDNs to fit your application at the time of purchase.
  • Before you begin testing with the CDN you will need to establish a calibrated drive levels corresponding to the test levels.
  • When used with compatible signal generator and spectrum analyzer, the CSAT test software can fully automate this calibration and also perform the test. Com-Power CDNs are calibrated to meet the latest CISPR 16 test standards.
  • For IEC 61000-4-6 Testing
  • RF Power Amplifiers
  • Wide Selection of CDNs
  • Software Automation
  • Directional Couplers
  • Attenuators

This system includes:

  • ACS-250-100 W Power Amplifier
  • CDNs (to be determined at the time of purchase)
  • ATTEN-6-100W
  • ATTEN-30-100W
  • DCU-1000-100W
  • Cable set
  • Appropriate CDN shorting adapters
  • ADA-515 150 Ohm to 50 Ohm adapters
  • 50 Ohm terminator

COSAT/CSAT Software (optional) at additional cost.