LISN 100 Amps CISPR 16 & ANSI C63.4

LI-1100C

150 kHz to 30 MHz | CISPR 16-1-2 / ANSI C63.4 | 50 microhenry | 100 Amps

line impedance stabilization network

LISN 100 Amps CISPR 16 & ANSI C63.4

  • High-current 100 A single-phase LISN, 150 kHz to 30 MHz — the highest current rating in the 1x series and the standard choice for industrial, three-phase commercial, and heavy-duty single-phase products requiring CISPR 16-1-2 / ANSI C63.4 conducted emissions testing.
  • Separately housed single-conductor networks with SUPERCON® shrouded sockets — one 50 µH air-core network per current-carrying conductor allows flexible installation in single-phase, dual-phase, DC, and three-phase configurations (a second pair extends to Wye / Delta three-phase).
  • Air-core 50 µH inductors, 270 VAC / 380 VDC, 50 Ω N-type RF port — CISPR 16-1-2 V-network performance maintained at 100 A continuous; insertion loss below 0.5 dB and isolation above 40 dB across 150 kHz to 30 MHz.
  • Industrial machinery and motor drives drawing up to 100 A — large variable frequency drives, industrial automation cabinets, machine tools, and CNC equipment requiring CISPR 11 / IEC 61000-6-4 industrial emissions qualification.
  • Three-phase commercial equipment via two LI-1100C pairs — data center UPS systems up to 25 kW, large commercial HVAC and refrigeration plants, three-phase server racks, and other commercial three-phase equipment qualifying for CE and FCC.
  • EV charging stations and high-power power conversion — AC charging infrastructure on single-phase 240 V circuits up to 100 A, large industrial battery chargers, and grid-tied inverters where the 100 A headroom is required.
  • Welding equipment and high-current industrial heating — arc welders, induction heaters, and resistive heating equipment drawing tens of amps continuous on single-phase mains with FCC Part 18 or CISPR 11 conducted emissions requirements.
  • NIST-traceable individual calibration with three-year warranty — impedance, phase, isolation, and insertion loss data supplied per network with calibration certificate; ISO 17025 accredited calibration available on request; recommended LIT-153A or LIT-930A transient limiter accessory available separately for EMI receiver input protection during high-current testing.

Features

  • Highest-current 1x Series LISN, 100 A AC with forced-air cooling — the flagship of the 1x Series for full-compliance testing of high-current industrial and 3-phase machinery.
  • 150 kHz to 30 MHz frequency range — covers the full CISPR Band B / commercial conducted emissions band for FCC Part 15, CISPR 22 / 32, EN 55022 / 55032 product certification testing.
  • Fully compliant with CISPR 16-1-2 and ANSI C63.4 — the 50 Ω / 50 µH V-network is calibrated to both standards, accepted by accredited EMC test laboratories worldwide.
  • 100 A AC / 70 A DC current rating — covers high-power industrial motors, welding equipment, large UPS systems, heavy-duty inverters, large EV chargers, and 3-phase machinery (when paired with a second unit).
  • High voltage rating: 440 VAC / 620 VDC — supports 480 V industrial 3-phase systems and high-voltage DC bus configurations beyond the 270 V / 380 V rating of the lower-current 1x Series models.
  • Forced-air cooling for continuous high-current operation — integrated cooling fans dissipate heat from the air-core inductor and current paths during extended testing of 100 A loads.
  • 50 µH air-core inductors prevent saturation — air-core construction keeps inductance constant at all currents up to the 100 A rating, ensuring the LISN impedance and emissions measurement remain calibrated.
  • Superior Electric SUPERCON® shrouded sockets — safe shrouded design rated for high-current connections; matching color-coded plugs included for mains and EUT wiring.
  • 50 Ω N-type female RF connector — standard interface for direct connection to EMI receivers and spectrum analyzers; insertion loss < 0.5 dB and isolation > 40 dB across 150 kHz to 30 MHz.
  • External transient limiter recommended — pair with the Com-Power LIT-153A or similar for impedance matching, out-of-band filtering, and EMI receiver input protection from the substantial transients of high-current EUTs.
  • 3-phase ready — a second LI-1100C pair extends the setup to 3-phase Wye or Delta industrial power systems.
  • Individually NIST-traceable calibrated — impedance, phase, isolation, and insertion loss data shipped with each unit. ISO 17025 accredited calibration available on request.
  • Three-year standard warranty — backed by manufacturer support.

Specifications

Model LI-1100C
Series 1x Series — Single-conductor pair, separately housed, with forced-air cooling
Specification CISPR 16-1-2 / ANSI C63.4
Application High-current Power Line conducted emissions tests
Frequency Range 150 kHz – 30 MHz
Inductors 50 µH (air-core)
Current Rating 100 Amperes AC (forced-air cooling), 70 Amperes DC
Voltage Rating 440 VAC (Line to Ground), 620 VDC
Cooling Forced-air (integrated cooling fans)
RF Connector 50 Ohms N-type (female)
Insertion Loss < 0.5 dB (150 kHz – 30 MHz)
Isolation > 40 dB (150 kHz – 30 MHz)
Mains & EUT Connections Superior Electric SUPERCON® shrouded sockets
Calibration Individual, NIST-traceable; ISO 17025 accredited available on request
Warranty Three-year standard warranty

All values are typical, unless specified. Ships as a pair of separately housed networks. All specifications subject to change without notice.

chart chart

Title Link
LI-1100C Datasheet View PDF
LI-1100C Manual View PDF

🔍 Not Sure This Is The Right LISN?

Compare all 25 Com-Power LISN models side-by-side with our interactive selection tool. Filter by current, voltage, frequency, and standards to find your perfect match.

Compare All LISN Models →

LI-1100C CISPR 16-1-2 / ANSI C63.4 High-Current Single-Phase LISN FAQs

What is the LI-1100C and what type of testing is it designed for?

The LI-1100C is a single-conductor, 50Ω, 50 μH Line Impedance Stabilization Network (LISN) designed for power-line conducted emissions compliance testing per CISPR 16-1-2 and ANSI C63.4 at currents up to 100 A AC with forced-air cooling. It is the highest-current model in the Com-Power single-phase CISPR 16-1-2 / ANSI C63.4 LISN range, designed for the most demanding commercial and industrial products including high-power motor drives, large UPS systems, high-power EV charging infrastructure, industrial welding equipment, large photovoltaic inverters, and any commercial product drawing up to 100 A.

What standards does the LI-1100C comply with?

The LI-1100C is fully compliant with CISPR 16-1-2 (specifying LISN electrical characteristics for conducted emissions measurements) and ANSI C63.4 (the American National Standard for FCC compliance testing). It is accepted by accredited EMC test laboratories worldwide for commercial product conducted emissions certification including CE marking under EN 55032, FCC Part 15, and other national standards derived from CISPR 16-1-2.

What is the frequency range of the LI-1100C?

The LI-1100C covers 150 kHz to 30 MHz — the same range as the LI-125C (25 A) and LI-150C (50 A). All three models share identical frequency coverage, RF connector type, 50 μH inductor value, and compliance standards. The only differences are maximum current rating and physical size required for the larger inductors and conductors at 100 A.

What is the current rating of the LI-1100C and what class of high-power products does it support?

The LI-1100C is rated at 100 A AC and 70 A DC — the highest in the Com-Power single-phase CISPR 16-1-2 range. This supports large industrial motor drives, high-power EV Level 2 and Level 3 charging stations, large UPS systems for data centers, industrial welding systems, large photovoltaic inverters and energy storage systems, high-power commercial HVAC systems, and high-power data center PDUs drawing up to 100 A single-phase.

What is the voltage rating of the LI-1100C and what power systems does it support?

The LI-1100C is rated at 440 VAC line-to-ground at 50–60 Hz and 620 VDC. The elevated 440 VAC rating (compared to 270 VAC for LI-125C/LI-150C) supports 277 VAC single-phase from 480 VAC North American industrial three-phase Wye systems, 380–415 VAC single-phase from European 400 VAC industrial three-phase Wye systems, and 620 VDC for EV battery storage, industrial DC bus, and high-voltage power conversion equipment.

Why does the LI-1100C require forced-air cooling and how does the system operate?

At 100 A, resistive losses in the air-core inductors generate substantial heat that passive convection cannot manage during extended compliance test sequences. The LI-1100C includes a fan powered by a dedicated adapter (120/230 VAC input, 6 VDC / 0.5 A output). The cooling fan must be running before the EUT is energized at or near full rated current, and must continue throughout the test. Operating at 100 A without the fan risks overheating the inductor insulation, damaging the unit and potentially altering inductance values.

How does the LI-1100C achieve its defined impedance and why are air-core inductors used at 100 A?

A 50 μH air-core inductor in series with the power line conductor, combined with capacitor and resistor networks, produces the defined 50 Ω CISPR impedance. Air-core inductors are essential at 100 A — any ferromagnetic core would saturate, causing inductance to collapse below 50 μH and invalidating the LISN impedance profile. Air-core inductors maintain a constant 50 μH at all currents up to the 100 A rating.

The LI-1100C is sold as a pair. Why is a pair required?

CISPR 16-1-2 and ANSI C63.4 require a LISN in series with each current-carrying conductor simultaneously (Line and Neutral). One LI-1100C is installed in the Line conductor and one in the Neutral. Measurements are performed one conductor at a time; the inactive RF port is terminated at 50 Ω. For three-phase systems, a second LI-1100C pair accommodates Phase B, Phase C, and Neutral.

What connectors does the LI-1100C use?

The LI-1100C uses Superior Electric SUPERCON® shrouded sockets at both mains and EUT ports, with color-coded plugs included. At 100 A and up to 440 VAC / 620 VDC, the shrouded connector design is critical for personnel safety. All power cables must be appropriately sized and rated for 100 A continuous operation.

Why is the mounting plate of the LI-1100C left unpainted and what grounding is required?

The unpainted mounting plate enables a low-impedance metal-to-metal bond to the test setup ground plane. At 100 A and up to 440 VAC, leakage currents through internal capacitors are potentially life-threatening if the earth bond is absent. The mounting plate must be bonded with a verified, direct low-impedance connection before any power connections are made. The test facility ground infrastructure must be verified as adequate for expected leakage current levels.

What RF connector does the LI-1100C use?

The LI-1100C uses a 50 Ω N-type female RF connector, connecting via 50 Ω coaxial cable to an EMI receiver or spectrum analyzer. The inactive LISN RF port must be terminated at 50 Ω. Com-Power recommends a Transient Limiter — at 100 A, large motor drives and industrial power converters generate much larger transient energy than lower-power products, making receiver protection especially important.

What insertion loss and isolation does the LI-1100C provide?

The LI-1100C provides insertion loss below 0.5 dB and isolation above 40 dB across 150 kHz to 30 MHz — identical to the LI-125C and LI-150C. Low insertion loss ensures measured voltages accurately represent actual EUT emissions. High isolation prevents facility power line noise from contaminating EUT measurement results at the RF port.

How is the LI-1100C individually calibrated and what calibration data is supplied?

Every LI-1100C is individually calibrated per CISPR 16-1-2 and ANSI C63.4. Impedance, phase, isolation, and insertion loss data for the specific unit are supplied with a certificate of calibration, providing precise insertion loss corrections for compliance testing and accounting for unit-to-unit component variations.

What is the warranty coverage for the LI-1100C?

The LI-1100C carries a three-year warranty from Com-Power Corporation. In addition to periodic recalibration on a one- or two-year interval, periodic inspection of the cooling fan to ensure it remains functional and that airflow is not obstructed is recommended.

What are the physical dimensions and weight of the LI-1100C?

Each LI-1100C network measures 10 x 10 x 21 inches (25.4 x 25.4 x 53.3 cm) and weighs 17 lbs (7.7 kg). The LI-1100C is substantially larger and heavier than the LI-125C and LI-150C. A complete single-phase installation requires significant floor or heavy bench space, 100 A-rated high-current wiring to the mains and EUT ports, and a powered cooling fan adapter running before EUT energization.

What is a Transient Limiter and why does Com-Power recommend one with the LI-1100C?

A Transient Limiter between the LI-1100C RF port and the EMI receiver provides impedance matching, attenuates out-of-band emissions below 150 kHz and above 30 MHz, and clamps voltage transients. At 100 A, large motor drives, UPS systems, and industrial power converters generate substantially larger transient energy than lower-power products, making the Transient Limiter especially important.

How does the LI-1100C complete the Com-Power single-phase CISPR 16-1-2 / ANSI C63.4 LISN range?

The LI-125C (25 A AC), LI-150C (50 A AC), and LI-1100C (100 A AC, forced-air) form the complete single-conductor CISPR 16-1-2 / ANSI C63.4 range, all sharing 150 kHz to 30 MHz, 50 Ω N-type RF, 50 μH air-core inductors, and individual calibration. Select the lowest-rated model accommodating the EUT current. The LI-1100C is required only when the EUT draws more than 50 A. Using a lower-current model at excessive current risks thermal damage and invalid measurements.

What types of equipment cannot be tested with the LI-1100C?

The LI-1100C is not suitable for: DO-160 or MIL-STD-461 testing (which require different network topologies); automotive CISPR 25 testing (use LI-550C); DC bus voltages exceeding 620 VDC; or three-phase systems without a second pair. At 100 A and up to 440 VAC, strict high-current and high-voltage safety procedures are mandatory — verified earth bonding, appropriately rated PPE, confirmed ground fault protection, and facility safety protocol compliance.

How is the LI-1100C used in conducted emissions compliance testing of a large industrial motor drive?

Large industrial motor drives drawing up to 100 A from single-phase 240 VAC or 277 VAC are installed through the LI-1100C pair between the AC mains and the drive’s AC input, with the drive output connected to the motor at representative load. Cooling fans are started first, then the drive is energized to rated input current. CE measurements from 150 kHz to 30 MHz on both Line and Neutral are compared against EN 55011 Class A industrial limits. The 100 A continuous rating with forced cooling allows full-rated-current testing throughout the entire CE scan duration — which may be 30+ minutes for a complete quasi-peak and average sweep — without thermally derating the LISN.

What does a conducted emissions test setup look like using the LI-1100C for a high-power EV charging station?

High-power Level 2 EV charging stations can draw up to 100 A from single-phase or split-phase 240 VAC. The LI-1100C pair connects between the regulated AC supply and the charger’s AC input; the charger output connects to an EV battery simulator at full rated charging current. Cooling fans are activated before the charger is energized. CE measurements from 150 kHz to 30 MHz capture switching frequency harmonics and rectifier noise that could interfere with nearby AM and FM receivers. The 440 VAC and 100 A ratings enable full-rated-current testing at both 120 VAC and 240 VAC charger configurations without changing the LISN.

How does an accredited EMC laboratory use the LI-1100C for third-party EN 55032 compliance testing of a high-power industrial power supply?

The laboratory confirms rated input current is up to 100 A, sets up the LI-1100C pair on the calibrated ground plane per EN 55032 / CISPR 16-1-2, and activates the cooling fans before energizing the supply at rated input current via a calibrated electronic load. Line and Neutral CE measurements from 150 kHz to 30 MHz are recorded with CISPR quasi-peak and average detectors. The LI-1100C’s individual calibration insertion loss correction is applied, and corrected results are compared against EN 55032 Class A or Class B limits. The LI-1100C’s calibration certificate, current rating, and CISPR 16-1-2 compliance are all documented in the formal test report.

How is the LI-1100C applied in conducted emissions testing of a large photovoltaic inverter for grid connection compliance?

Grid-tied PV inverters for commercial and industrial solar installations may output up to 100 A at single-phase 240 VAC. The LI-1100C pair is installed between the inverter AC output and a grid simulator or bidirectional power supply configured to absorb full rated output power. CE measurements from 150 kHz to 30 MHz at full rated output power per EN 55011 or FCC Part 15 capture switching frequency harmonics and intermodulation products generated by the DC-to-AC conversion stage that could appear on the utility grid and interfere with grid communication systems and other grid-connected equipment.